STANDARD DIMENSIONS

Figure 11-1. Standard Dimensions

Figure 11-2. Panel Cut-Out

DISASSEMBLY
You may need to open up the meter to check or change the 115 or 230 Vac power jumpers. You may also need to connect or disconnect serial communication boards. To remove and access the main board, follow the steps described in Section 3.3.

GENERAL INFORMATION
Programmable Functions
Rate (frequency) and totalizer

Scale Factor
-99999 to 999999 with a choice of 6 decimal point positions (9.9.9.9.9.9.), multiply or divide

Offset

-99999 to 999999 with a choice of 6 decimal point positions (9.9.9.9.9.9.)


INPUT CHARACTERISTICS
Maximum Frequencies Inputs - Square Wave, (0 to +5 V)
  • 30 kHz: S1-5 open
  • S1-1 and S1-4 closed


  • Maximum Input Voltage Signal
    S1-6 and S1-7 open 60 V rms, input protected to 120 V rms; S1-7 closed: 18 V rms S1-6 closed: 15 V rms

    Maximum Input Voltage (all other inputs)
    0 to 5 V

    Minimum Signal
    25 mV, rms (S1-3 closed)

    Trigger Slopes
    Positive or Negative, select by S1-8

    External Gate and Reset Levels
    Low level = less than 0.8V
    High level = more than 3.5V

    Input Impedance - Signal Input (S1-6 and 7 open)
    243K pull up to excitation voltage if signal is between 0 and excitation voltage.

    External Gate and Reset Inputs
    20k pull up to +5 V


    DISPLAY

    Type
    6 digit, 7-segment, LED

    Display Digit Height

    14.2 mm (.56 in) Symbols: -.8.8.8.8.8. and 8.8.8.8.8.8.

    Decimal Point

    6 positions; programmable; Fixed or auto-range

    Leading Zeros: Blank

    Overflow Indication

    Shows in rate or total mode; in exponential format to 9.99 E9

    Update Rate
    3 readings per second (unless modified by RS-232)

    Displayed Value

    (Measurement ╖ Scale Factor) + Offset 11.6

    TIME BASE
    Internal Clock Reference
    921.6 KHz

    Stability:
    ▒2 ppm/░C

    Fine-Calibration Method

    Programmable calibration value accessible through RS232 communications. Requires frequency calibrator to calibrate the meter

    Calibration accuracy at 25░C:
    ▒10 ppm


    RATE FUNCTION
    Frequency range
    0.5 Hz to 30 KHz. The meter utilizes exponential format above 6 digits. 9.99 E9 signifies maximum display.

    Unit of Measurement

    Rate: Hertz (Scale Factor = 1)

    Accuracy at 25░C

    ▒1 LSD ▒0.01% of the input (0-5 V (square wave) input)


    TOTALIZER
    Count direction
  • Up or Down (determined by sign of scale factor)


  • Internal count capacity

  • (500,000,000,000) input pulses. The display utilizes exponential format above 6 digits.
  • 9.99 E9 signifies maximum display.


  • AC power failure provisions
  • You may store the display value in nonvolatile memory and restore with power (in totalize mode only).


  • ANALOG OUTPUT OPTION BOARD
    Max Load Impedance for Current Output: 500 ohm
    Resolution: 15 bits
    Accuracy, 0-10V:99.8%
    Accuracy, 4-20 mA:
    99.75%

    Response Time
    Total Mode: 80 mSec (99% of final value)
    Rate Mode:
    360 mSec (99% of final value)


    RELAY OUTPUT OPTION BOARD
    2 Form "C" on/off relays. Configurable for latched and unlatched by software.
    Max Current: 5Amps, resistive load.
    Max Voltage: 250Vac or 28Vdc


    POWER
    AC voltage (standard): 115 or 230 Vac ▒15%
    Frequency: 49 to 63 Hz
    Power consumption: 11.5 W, max.

    External Fuse Protection Recommended
    AC 115V: IEC 127-2/III, 100mA, 250V (T Type) or UL Slow-Blow 125mA, 250V
    AC 230V: IEC 127-2/III, 50mA, 250V (T Type) or UL Slow-Blow 1/16 Amp, 250V

    Excitation Output (Select by S1-1 and S1-2)

  • 5 V @ 50 mA maximum
  • 8.2 V @ 70 mA maximum
  • 12.4 V @ 100 mA maximum
  • For excitation currents more than 50 mA, the ac voltage must be 115 or 230 Vac ▒10%.


  • DC Power/Battery backup
  • 7.5-13 V, 30 mA (typical), without any option board. The display goes blank, but the meter continues to totalize.
  • 260 mA is required to maintain the front-panel digital display. Add 180 mA for analog output board, 120 mA for relay board, 20 mA for RS-232 board.


  • ENVIRONMENT
    Operating Temperature: 0░ to 50░C (32░ to 122░F)
    Storage Temperature: -40░ to 85░C (-40░ to 185░ F)
    Relative Humidity: 90% at 40░C (non-condensing)

    MECHANICAL
    Panel cutout: 1/8 DIN 3.62" x 1.78" (45 x 92mm)
    Weight: 1.27 lb (575 g)
    Case material: Polycarbonate, 94 V-O UL rated